Technical
Data
TYMPANOMETRY
MODE
Probe
Frequency .
. . . . .226 Hz +/- 3%
Probe
Intensity .
. . . . . .85 dB SPL, +/- 2.0 dB
measured into a 2 cc coupler,
AGC
controlled
Probe
Distortion .
. . . . .less than 5%
Pressure
Range . .
. . . . .+200 to -400 daPa. Sweep stops -
200
daPa from the peak
Pressure
Accuracy .
. . . .+/- 15% or +/- 10 daPa,
whichever is greater
Pressure
Sweep Rate .
. .400 daPa/sec nomina
EST
Time of Test .
. . . . .approx. 5 seconds
Compliance
Range . .
. . .0.2 - 4.5 cc
Compliance
Accuracy . .+/- 0.1cc or 5%,
whichever is
greater over the Compliance Range
Ear
Canal Volume Range 0.2
- 6.0 cc
Uncompsensated
Compliance
. . . . . . . .5.0
cc
Compensated
Range . .
. .0.0 - 1.0cc, 0.0 - 2.0cc, 0.0 - 3.0cc
REFLEX
MODE
Reflex
Tone . . . . . . . . . .1
KHz
Reflex
Level . . . . . . . . . .100
dB HL
Accuracy
. . . . . . . . . . . .+/-
3%
Total
Harmonic
Distortion <5%
Determination
. . . . .
. . .Compliance change >.062 cc
Pressure
. . . . . . . . . . . . .Automatically
returns to peak
compliance point
DISPLAY
Type
. . . . . . . . . . . . . . . .LCD
PRINTER
Type
. . . . . . . . . . . . . . . .Fujitsu Thermal
PROBE
Weight
. . . . . . . . . . . . . .4 oz.
Dimensions
. . . . . . . . . . .1"
x 2-1/2" x 1-1/4"
MECHANICAL
Dimensions
. . . . . . . . . . .11-3/4"
x 8-1/2" x 5-1/2" envelope
Weight
. . . . . . . . . . . . . .5 lbs
Voltage
. . . . . . . . . . . . . .117 volt +/- 10%
ENVIRONMENTAL
Operational
Temp. .
. . .15° - +40° C
Storage
Temp. . .
. . . . . .0° to +60° C
Relative
Humidity .
. . . .5° to 90%
AUDIOMETRY
(Race Car Tympanometer
w/Audiometer only)
Frequencies
. . . . . . . . . .250,
500, 1000, 2000, 3000, 4000,
6000,
and 8000 Hz
Decibel
Range . .
. . . . . .0-85 dB in 5 dB steps
Frequency
Accuracy .
. .+/- 2%
Noise
Level . . . . . . . . . . .10
dB HL or 70 dB below the test
tone, whichever is greater
Distortion
. . . . . . . . . . . .Less
than 3% THD
Level
Accuracy .
. . . . . .+/- 3 dB of levels specified in
ANSI
S3.6-1996
TONE
PRESENTATION
Rise/Fall
Time . .
. . . . . .35 +/- 15 milliseconds
On/Off
Ration .
. . . . . . .greater than 70 dB at all
frequencies
|
OUTPUT
Earphones
. . . . . . . . . . .Telephonics THD-39 w/100 ohm
coils
Crosstalk
. . . . . . . . . . . .-70 dB or less than
- 5dB HL,
whichever is greater
CPT
CODES
92567
. . . . . . . . . . . . .Tympanometry
92568
. . . . . . . . . . . . .Acoustic Reflex
92552
. . . . . . . . . . . . .Pure Tone Threshold

Race Car, Quik
Tymp Series Pilot
Technical
Data
MECHANICAL
Dimensions
. . . . . . . . . . .12"
x 9" x 2" (30.5 x 23 x 5 cm)
Weight
. . . . . . . . . . . . . .2.59 lbs (1169.80 grams)
Power
Consumption .
. .108 volt - 132 volt AC, 50-60 Hz, 10 watts
ENVIRONMENTAL
Operational
Temp. .
. . .15° - +40° C
Storage
Temp. . .
. . . . . .0° to +60° C
Relative
Humidity .
. . . .5° to 90%
AUDIOMETRY
Frequencies
. . . . . . . . . .500,
1000, 2000, 4000, Hz (Pure Tone)
Decibel
Range . .
. . . . . .Voice - 50, 45, 40, 35, 30, 25, 20, 15 dB level
ranges
Pure
tone - 40, 30, 20 dB HL
Frequency
Accuracy .
. .+/- 3%
Noise
Level . . . . . . . . . . .10
dB HL or 70 dB below the test tone, whichever is greater
Distortion
. . . . . . . . . . . .Less
than 3% THD
Level
Accuracy .
. . . . . .+/- 3 dB of levels specified in ANSI S3.6-1996
Sound
Output Level .
. .40, 30, 20 dB HL +/- 3 dB
TONE
PRESENTATION
Rise/Fall
Time . .
. . . . . .35 +/- 15 milliseconds
On/Off
Ration .
. . . . . . .greater than 70 dB at all frequencies
OUTPUT
Earphones
. . . . . . . . . . .Telephonics THD-39 w/100 ohm coils
Crosstalk
. . . . . . . . . . . .-70 dB or less than
- 5dB HL, whichever is greater
CALIBRATION
In
order to ensure a safe and accurate measure for all your hearing
exams, it is strongly
recommended that units be calibrated on a yearly basis.
CPT
CODES
92583
. . . . . . . . . .Select Picture Audiometry
92551
. . . . . . . . . .Screening Audiometry
(Pure Tone).

|