Here is the complete procedure as you would perform
it in your practice:
Step 1: Preparing for the Test
Before beginning the procedure read the Product
Insert and put on gloves. This demo shows a fingerstick, but a
venous sample may also be used.
Remove a sample dilution pack and an A1cNow monitor
pack from the box. Ensure the monitor has been at room temperature
for at least 60 minutes before opening the monitor pouch.
Gather a lancet and gauze for use in collecting the
blood sample. Tear off a patient label and fill out with patient
name and date of testing.
Open both packages by grasping and tearing at the
notch. Place all items on a clean surface. Use the monitor within 15
minutes after removing it from the pouch. Place the monitor on a
flat surface away from heat or cold. Place the previously filled-out
patient label on the top of the monitor.
Remove the capillary & transfer pipettes, the
tube holder, and the sample dilution vial from the pack. Remove the
cap from the vial. Place the vial in the tube holder.