|

Background
A1cNow® is a single-use, disposable monitor for the
quantitative measurement of %A1c (A1C) in fingerstick or venous whole
blood samples. With one drop of blood and just three simple steps, the
test can be performed in a physician's office or by the patient at home.
Once diluted sample is added to the monitor, there are no further steps,
and results are available in about 8 minutes. The monitor performs over 25
internal chemical and electronic quality control checks with each test,
including checks for potential hardware or software errors, and potential
reagent strip errors. An error code is reported in place of a result if
any quality check does not pass.
The A1cNow test is certified by
the National Glycohemoglobin Standardization Program (NGSP) and
therefore traceable to the DCCT. Further, the test is factory calibrated
with a set of blood samples that have been quantitated by an NGSP
certified laboratory using an NGSP certified method. Since in-field
calibration is neither needed nor possible, the user cannot alter the
accuracy of the test. Allowing for the normal variation seen with all
laboratory methods, a 7% HbA1c on A1cNow will thus be, on average, the
same as a 7% HbA1c on other instruments that are also calibrated to NGSP
standards.
Accuracy
Independent Field Study Using A1cNow® (Pacific
Northwest)
November 21, 2002
A large health care organization in the Pacific Northwest
recently performed a correlation study using A1cNow. The laboratory took
78 EDTA venous samples, performed their routine laboratory testing, A1cNow
testing, and sent an aliquot of the sample to the National Glycohemoglobin
Standardization Program (NGSP) Secondary Reference Laboratory in Columbia,
Missouri. All results were returned to Metrika, and to the health care
organization.
Data analysis performed by Metrika showed the following.
Bias between A1cNow and the NGSP lab was negligible (0.00%HbA1c), showing
accuracy of over 99% for this study. Bias between the health care
organization laboratory and A1cNow was small, -0.25%HbA1c, and bias
between the health care organization laboratory and the NGSP laboratory
was also small, +0.31%HbA1c. Regression statistics are presented in the
table below.
| |
Slope |
Y-intercept |
"r" |
Average
Bias
%HbA1c |
|
A1cNow vs. NGSP |
1.02x |
-0.13 |
0.96 |
0.00 |
|
A1cNow vs. Lab |
0.93x |
+0.28 |
0.95 |
-0.25 |
|
Lab vs. NGSP |
1.10x |
-0.37 |
0.99 |
+0.31 |
Graphic representations of the data are shown,
below.
A1cNow vs. NGSP:

A1cNow® Comparison to Quest Diagnostics
Clinical Trials Laboratory
December 5, 2002
An in-house clinical study was performed at Metrika on
October 14, 2002. The objective of the study was to obtain comparative
data between A1cNow and Quest Diagnostics Clinical Trials Laboratory,
which uses the BioRad Variant II®. Fingerstick samples were collected for
A1cNow and Variant II testing. Venous samples were collected and sent to
the National Glycohemoglobin Standardization Program (NGSP) Secondary
Reference Laboratory in Columbia, Missouri.
Data analysis performed by Metrika showed the following.
Average bias between A1cNow and the NGSP lab was small (+0.1%HbA1c),
showing accuracy of approximately 98% for this study. Bias between
the Quest laboratory and A1cNow was also small, +0.2%HbA1c, as was the
bias between the Quest laboratory and the NGSP laboratory, at -0.1%HbA1c.
Means of duplicate pairs were plotted and least-squares linear regression
statistics were calculated. Regression statistics are presented in the
table below.
| |
Slope |
Y-intercept |
"r" |
Average
Bias
%HbA1c |
|
A1cNow vs. NGSP |
0.86x |
+1.08 |
0.99 |
+0.1 |
|
A1cNow vs. Quest |
0.91x |
+0.83 |
0.98 |
+0.2 |
|
Quest vs. NGSP |
0.94x |
+0.34 |
0.99 |
-0.1 |
Graphic representations of the data are shown
below.
A1cNow vs. NGSP

|